Show/Hide Toolbars

Irinos Measurement System

Navigation: MscDll Reference Guide > Opcodes and parameters > Opcodes: Initialization

opcRIV: Read inventory (number of boxes)

Scroll Prev Top Next More

This opcode allows reading the number of available Irinos-Boxes from the Irinos-System.

 

Overview


Opcode:              0x01

Name:                opcRIV

Function:            Read inventory (number of boxes)

Parameter type:        String

 

DLL-Function


MSC_WriteCommand

 

Request string to the Irinos-System


None. No data needs to be sent.

 

Response string from the Irinos-System


#{Number of Irinos-Boxes};{Number of modules}#

Number of Irinos-Boxes

Total number of Irinos-Boxes used in the Irinos-System.

Number of modules

This value is identical with "Number of Irinos-Boxes". It is only used for backward compatibility to older systems.

 

Example:

#3;3#     -> 3 Irinos-Boxes are used.

This opcode cannot return an error response string.

 

Comment


The Master-Box (with Ethernet interface to the PC) is also counted. The example above could be a combination of the following Irinos-Boxes:

1 Master-Box IR-MASTER-KB1-68-68-SYSP-ETHIL
1 Slave-Box for inductive probes IR-TFV-8-TESA-M16-IL
1 Slave-Box for incremental encoders IR-INC-4-SEL1VSS-D15F-IL