This opcode allows reading the statusword for dynamic measurement.

 

Overview


Opcode:              0x44

Name:                opcRSW

Function:            Read statusword (for dynamic measurement)

Parameter type:        Binary

 

DLL-Function


MSC_WriteCommand

 

Request data to the Irinos-System


None

 

Response data from the Irinos-System


Longword (32 Bit) with status information. See table below.

 

 

Comments


A status-change of the following bits may be delayed:

oTrigger X was active and is now inactive

oTrigger X: Min 1 trigger pulse occurred

oDynamic measurement X was active and is now inactive

oDynamic measurement X: Min 1 value has already been sampled

 

Statusword for dynamic measurement


Bit No.

Function

0

Trigger 1 active

1

Trigger 1 was active and is now inactive

2

Trigger 1: Min 1 trigger pulse occurred

3

 

4

Dynamic measurement 1 is active

5

Dynamic measurement 1 was active and is now inactive

6

Dynamic measurement 1: Min 1 value has already been sampled

7

Dynamic measurement 1: Reading values by the PC is active

8

Dynamic measurement 1: Internal sample buffer is full

9

 

10

 

11

 

12

 

13

 

14

 

15

 

16

Trigger 2 active

17

Trigger 2 was active and is now inactive

18

Trigger 2: Min 1 trigger pulse occurred

19

 

20

Dynamic measurement 2 is active

21

Dynamic measurement 2 was active and is now inactive

22

Dynamic measurement 2: Min 1 value has already been sampled

23

Dynamic measurement 2: Reading values by the PC is active

24

Dynamic measurement 2: Internal sample buffer is full

25

 

26

 

27

 

28

 

29

 

30

 

31